We use cookies to ensure that we give you the best experience on our website. If you click 'Accept all cookies' we'll assume that you are happy to receive all cookies and you won't see this message again. If you click 'Reject all non-essential cookies' only necessary cookies providing core functionality such as security, network management, and accessibility will be enabled. Click 'Find out more' for information on how to change your cookie settings.

AbstractPurposeTo optimize pseudo‐continuous arterial spin labeling (PCASL) parameters to maximize SNR efficiency for RF power constrained whole brain perfusion imaging at 7 T.MethodsWe used Bloch simulations of pulsatile laminar flow to optimize the PCASL parameters for maximum SNR efficiency, balancing labeling efficiency and total RF power. The optimization included adjusting the inter‐RF pulse spacing (TRPCASL), mean B1+ (B1+ave), slice‐selective gradient amplitude (Gmax), and mean gradient amplitude (Gave). In vivo data were acquired from six volunteers at 7 T to validate the optimized parameters. Dynamic B0‐shimming and flip angle adjustments were used to avoid needing to make the PCASL parameters robust to B0/B1+ variations.ResultsThe optimized PCASL parameters achieved a significant (3.3×) reduction in RF power while maintaining high labeling efficiency. This allowed for longer label durations and minimized deadtime, resulting in a 118% improvement in SNR efficiency in vivo compared to a previously proposed protocol. Additionally, the static tissue response was improved, reducing the required distance between labeling plane and imaging volume.ConclusionThese optimized PCASL parameters provide a robust and efficient approach for whole brain perfusion imaging at 7 T, with significant improvements in SNR efficiency and reduced specific absorption rate burden.

Original publication

DOI

10.1002/mrm.30527

Type

Journal article

Journal

Magnetic Resonance in Medicine

Publisher

Wiley

Publication Date

28/05/2025